Table 2: Alarm Default Parameters for Volume Ventilation for 10 kg to 70 kg
Table 3: Default Alarms Settings
Low Pressure 5 cmH
2
O
Apnea Time 20 Sec
High FiO
2
Set FiO
2
Value + 20% (Factory Setting 80%)
Low FiO
2
Set FiO
2
Value - 10% (Factory Setting 50%)
Leak (%) 100% (off)
Inverse I:E Ratio ON
Low V
T
Alarm Range 85%
Table 4: Default Parameters for Pressure Ventilation for 10 kg to 70 kg
Patient Weight RATE
P
Insp.
(above
PEEP)
I Time
Volume Limit
(kg) (lb) (bpm) (cmH
2
O) (sec) (mL)
10 22 30 20 0.6 200
15 33 25 20 0.8 300
20 44 20 20 1.0 400
30 66 18 20 1.1 600
40 88 16 20 1.2 800
50 110 14 20 1.4 1000
60 132 12 20 1.7 1100
70+ 154 12 20 1.7 1200
Alarm Paramete
Weight (kg)
10 15 20 30 40 50 60 70+
High Pressure (cmH
2
O) 23 25 30 33 35 40 40 40
High Rate (bpm) 40 35 30 30 30 30 30 30
Low Rate (bpm) 10 10 10 8 6 6 6 6
High Min Vol. (LPM) 4 5 7 10 12 14 15 16
Low Min Vol. (LPM) 1 1 1 1 2 2 2 2