EasyManuals Logo

GE iVent 201 Series User Manual

GE iVent 201 Series
290 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #168 background imageLoading...
Page #168 background image
156
Figure 145: The Low Tidal Volume Alarm Range Pop-Up Window
4. Turn the Control Knob to adjust the percentage between 0% (Off) and 85%.
5. Press the dial to confirm the settings.
6. You are returned to the Alarm Settings screen. Dial and select Accept to
accept the Alarm Options settings.
Disabling or Enabling the Patient Circuit Disconnect Alarm
When ventilating with Adaptive Bi-Level, under some circumstances it is
appropriate to turn off the Patient Circuit Disconnect Alarm.
WARNING The default setting for the Patient Circuit Disconnect Alarm is ON. It
should be shut off under medical discretion ONLY if necessary. Be sure
that you are operating in accordance with facility policy and patient
needs.
NOTE This option is available in Adaptive Bi-Level mode only.
To disable the patient disconnect alarm:
1. From the Main screen select Menu – Alarm Settings - Options. The Alarm
Options screen is displayed.
2. Dial the Control Knob until Patient Circuit Disconnect is selected. The state
of the Alarm (ON or OFF) is shown in the right column of the screen.
3. Press the Control Knob. An ON/OFF selection pop up window appears
(Figure 146).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE iVent 201 Series and is the answer not in the manual?

GE iVent 201 Series Specifications

General IconGeneral
BrandGE
ModeliVent 201 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals