1 Introduction 
25 
 
INTENDED USE 
The iVent™201 MR Conditional is a portable, computer controlled, electrically 
powered intensive care ventilator intended to provide continuous or intermittent 
ventilatory support for the care of individuals who require mechanical ventilation. 
Specifically, the ventilator is applicable for use with adult through pediatric 
patients, who require invasive or non-invasive assistance via the following 
general modes of ventilatory support, as prescribed by an attending physician: 
  Assist/Control (Pressure Controlled or Volume Controlled) 
  SIMV (Pressure Controlled or Volume Controlled) 
  CPAP/PSV 
The iVent™201 MR Conditional ventilator is suitable for use in the ICU and all 
other hospital areas, including Magnetic Resonance (MR) environment, not to 
exceed a 3.0 Tesla static magnetic field, in all hospital-type facilities, alternate 
care sites, transport, emergency and in the home environment.  The iVent™201 
MR Conditional ventilator is MR conditional.  
The optional non-invasive Pulse Oximeter is intended for non-invasive monitoring 
of oxygen saturation and pulse rate and is suitable for use in all above mentioned 
areas, excluding MR environments. 
The iVent™201 MR Conditional ventilator is a restricted medical device intended 
for use by qualified, trained personnel under the direct supervision of a physician. 
 
USE OF THE IVENT
TM
201
 
WITH MRI 
The iVent™201 ventilator (serial number 15 000 and higher) is classified as MR 
conditional for 1.5T and 3T MR scanners. This means that the iVent
TM
201 is safe to 
use in the MR environment if the operation conditions specified in this section are 
met. 
For the MRI conditional specifications refer to MRI Conditional Tests 
Specifications, on page 29 of this document. 
 
CAUTION  The iVent
TM
201 ventilator should not be subjected to field strength 
greater than 100 Gauss and must be kept outside of 100 Gauss 
perimeter. 
 
NOTE The iVent
TM
201 MR Conditional has been demonstrated to pose no known 
hazards in MR Environment under specified conditions of use. However, MR 
image quality may degrade when working with the ventilator in a specific MR 
suite. It is recommended to perform a pre-qualification image quality test to 
ensure an acceptable level of image quality. For more details contact your 
local sales representative