EasyManua.ls Logo

GE iVent 201 Series - O2 Calibration

GE iVent 201 Series
290 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Care, Maintenance and Tests
177
For a complete description of the Calibration procedure refer to VersaMed Service
Manual.
O2 Calibration
It is recommended to perform O2 calibration every 3 months to ensure system
integrity. The calibration process comprises two separate calibration procedures,
one for 100% oxygen and another for 21% oxygen.
NOTE You are required to perform O2 calibration every 6 months
To calibrate the O2:
1. Turn the control-knob on the Main screen to select Menu –
Maintenance. A caution pop up window appears, which warns that the
menu is a restricted maintenance area, and is for service personnel only.
2. Select Yes. The Maintenance screen is displayed (Figure 151).
Figure 151: Maintenance Screen
3. Select Calibration. The Calibration screen is displayed (Figure 152).
Figure 152: Calibration Screen
4. Select Calibrate O2 System.

Table of Contents

Related product manuals