EasyManua.ls Logo

GE iVent 201 Series - Cleaning and Sterilization

GE iVent 201 Series
290 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
248
Humidifier to ensure adequate drainage to it.
The Y-transducer should always slope slightly towards the Water trap (greater
than 10
o
) to ensure adequate drainage to the Water trap.
Figure 178: iVent
TM
201 Reusable Patient Breathing Circuit
Figure 179: Exhalation Valve components
Figure 180: Example of water trap tubing assembly
Cleaning and Sterilization
Disassembly
1. Detach the one-way valve from the Y-Connector and the long (inspiratory)
corrugated tube. Note: Do not remove the tube holders that support the
sensor tubes from the corrugated tube.
2. Disassemble the exhalation valve as follows: open the top cover together
with its control tube, and GENTLY remove the internal membrane. (Note: do
A
-
Ventilator outlet
B -
Flow sensor
C-
External exhalation valve
c
D- External exhalat
i
E- Flow sensor
F - Wye connector
G- Nebulizer outlet

Table of Contents

Related product manuals