16
ENVIRONMENTAL SPECIFICATIONS
Parameter
Operating Temperature 0 to +50 °C / +32 to +122 °F
Storage Temperature Without battery: -15 to +70 °C / +5 to +158
°F
With battery: -15 to +30 °C / +5 to +86 °F
Relative Humidity 15 to 95% @ 30 °C / 86 °F
Water and Dust
Resistance
IP-54 (Splash proof)
Atmospheric Pressure 430 - 825 mmHg (up to 15,000 feet)
Vibration IEC 68-2-6, IEC 68-2-34
MIL-STD-810E
Shock IEC 68-2-27 (100 g)
MIL-STD-810E
Total External Sound
Level
40 – 45 dBa at one meter
POWER SUPPLY
Parameter
External AC 100 to 240 V, 50-60 Hz, max 1.6 A.
External DC Battery 12 to 15 V max 8.5 A.
Internal Battery Sealed Lead-Acid, 12 V 8 Ah (rechargeable)
Recharge Time Over 8 hours, <1 Ampere
Operating Time (internal
battery)
Up to 2 hours (varies with ventilation
parameters)
Extended Battery Sealed Lead-Acid, 12 V, 12 Ah (rechargeable)
Extended Battery
recharge time:
Over 24 hours, <1 Ampere
Operating Time (extended
battery)
Up to 4 hours (varies with ventilation
parameters)
Operating Time (external
battery)
Up to 8 hours (varies with ventilation
parameters)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE iVent 201 Series and is the answer not in the manual?

GE iVent 201 Series Specifications

General IconGeneral
BrandGE
ModeliVent 201 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals