230
Figure 173: Open Loop Warning Message
If Standby was selected, the following Confirmation message pop-up window
appears (Figure 174):
Figure 174: Standby Confirmation Message
The Open Loop ventilation is an approximation of the Volume Controlled
ventilation with pressure limit. Ventilation is based on the average measured
inhaled volume similar to apnea ventilation. See Table 16 above. Pressure is
measured via an internal pressure sensor.
Upon the restoration of normal conditions, the unit automatically restores the
previous ventilation parameters after 2-3 proper breaths are detected. The user is
prompted to restore the previous ventilation parameters manually at any time or
to select the Standby Options.
ADAPTIVE BI-LEVEL MODE
See Chapter 5.