3 Operating the iVentTM201 – Setting Modes and Parameters 
67 
 
The chosen values are not operational until you confirm your selection by 
pressing the Control Knob, which takes you back to the Main screen.  The new 
value(s) you have selected are shown in black. 
 
NOTE  To abort the selection process and return the parameter to the previously 
chosen value, press the Clear button. 
To return the selection of ALL ventilation parameters to their default levels for 
a given patient weight see page 105. 
 
ADJUSTING THE BREATH RATE 
The following procedures explain how to adjust the breath rate, which is 
measured in breaths per minute. 
To adjust the breath rate during the ventilation: 
1. From the top most left side of the Main screen select the Respiratory Rate 
option (Figure 35). 
 
Figure 35: Respiratory Rate Selected 
Or select the Rate setting option from the Mode Parameter screen (Figure 36).