Appendix D: Theory of Breath Delivery 
215 
 
  Pressure control breath - mandatory breath in which the ventilator provides 
constant pressure at a preset level during inspiration. 
  Patient Breath. Any breath in which the patient initiates and terminates the 
inspiration phase. This breath type is divided into two secondary types 
according to the method by which the inspiratory phase is controlled: 
o  Spontaneous breath - the ventilator helps the patient to breathe by 
maintaining the inspiratory pressure at the baseline level (PEEP). 
o  Pressure Support breath - the ventilator elevates the inspiratory 
pressure to the preset support pressure level above the baseline and 
maintains this pressure during the inspiration phase. 
ADAPTIVE FLOW AND ADAPTIVE I-TIME 
Unique to the iVent
TM
201, Adaptive Flow
TM
 and Adaptive I-Time™ are patient 
responsive controls to establish appropriate peak flow and inspiratory time 
during volume control breath delivery. The main goal of the algorithm is to match 
the patient demand for flow during inhalation while reducing airflow hunger and 
increasing patient comfort.  
Should the Adaptive Flow algorithm determine that the peak flows achieved by 
the patient’s spontaneous breathing exceed the peak flows that are determined 
necessary to achieve the set tidal volume, then the algorithm matches the 
patient’s peak flow so as to avoid the feeling of “air hunger” in the patient. In this 
situation, the ventilator does not realize an I:E ratio of 1:2. 
 
In the absence of sufficient spontaneous patient efforts, Adaptive Flow and 
Adaptive I-Time work together in volume control modes (SIMV and A/C). When 
used together (The default state for SIMV Volume Control mode) these two 
features seek to achieve an I:E ratio of 1:2 (As close as possible within clinical 
limitations, such as Pressure Limit and Alarm Pressure Cutoff settings as well as 
minimum and maximum Inspiratory Time allowed). 
Changes in the overall breath rate are tracked and the Adaptive I-Time algorithm 
adjusts the inspiratory time over approximately 10 breaths in order to maintain 
the I:E ratio at 1:2. The Adaptive Flow algorithm accommodates changes in the i-
time and automatically adjusts the peak flow so that the delivery of the set tidal 
volume for the i-time determined by the Adaptive I-Time algorithm is assured. 
 
  The Adaptive I-time changes inspiratory time as required to achieve a 1:2 I:E 
ratio. 
  If respiratory rate increases, the inspiratory time decreases and the 
mandatory peak inspiratory flow increases to deliver the set tidal volume. 
  If respiratory rate decreases, the inspiratory time increases and the 
mandatory peak inspiratory flow decreases to deliver the set tidal volume. 
If flow is insufficient to deliver the tidal volume, inspiratory time gradually 
increases in an attempt to deliver the set tidal volume. In this situation the 
ventilator does also not achieve I:E ratios of 1:2. For this reason, the user is 
advised to leave the inverse I:E ratio alarm "ON" when using Adaptive Flow and 
adaptive I-Time. If inverse I:E ratios are achieved, it is recommended that the 
ventilator controls are set manually.