PERFORMANCE PARAMETERS 
Ventilation Component  Parameters/ 
Range 
 
Unit of 
Measure 
Accuracy 
Set 
 
Displayed 
Page 
 
Respiratory rate  1-12 
12-80 
bpm 
1 
2 
Y Y 67 
Inspired Tidal Volume 
Measurement 
50-2000 mL 
10% 
 
Y N 69 
Exhaled Tidal Volume 
Measurement 
50-300 
 
300-2000 
mL 
10 mL or 
+15% 
whichever is 
greater 
+10%  
N Y  NA 
Inspiratory Pressure Limit  5 to 80  cmH2O 
5 
Y Y 71 
Inspiratory Time  0.2 to 3 or 
Adaptive I-
Time 
sec. 
10% 
Y Y 80 
Peak Flow  up to 120 or 
Adaptive 
Flow 
L/min 
10% 
Y Y 74 
Peak Flow (Spontaneous)  up to 180  L/min 
10% 
Y Y 74 
Oxygen Mix (FiO2)  21 to 100%   
10% of the 
set value or 
3% absolute 
whichever is 
greater 
 
Y Y 73 
PEEP  0 to 40  cmH2O 
1 or 10%, 
whichever is 
greater 
Y Y 78 
Trigger sensitivity   1 to 20, off 
-0.5 to –20,  
off  
L/min 
cmH2O 
 
 
Y Y 79 
PSV  0 to 60  cmH
2
O 
 10% 
Y Y 76 
Positive Pressure relief 
valve 
80 cmH
2
O    N N NA 
Controlled Pressure  5 to 80  cmH
2
O 
5 
Y Y NA 
Flow Termination  10-90  %  +10% 
Y Y 84 
Rise Time  Mid, High, 
Max, Auto 
 NA Y Y 82 
FiO
2 
at power up  21%, 40%, 
60%, 100% 
(selectable) 
  Y Y 185