2 Setting up
35
provided by external batteries may change. To prevent switch-offs from
the external battery due to normal fluctuations in DC voltages, the
iVent
TM
201 is programmed to average the external DC voltage reading
over time. Therefore, if the user disconnects the external battery, the
ventilator interprets this as a dying battery and switches off the external
DC source for power. To re-enable input from the external battery
source, the user must leave the external battery disconnected for 40
seconds before reconnecting the external battery
INTERNAL BATTERY
The ventilator switches to an internal battery when it detects a loss of external
power. When fully charged, the internal battery can supply approximately 1 to 2
hours of power, under typical conditions and settings.
NOTE The extended internal battery can supply approximately 3 to 4 hours of
power under typical conditions and settings
CAUTION The iVent
TM
201 should only be used with a properly functioning battery.
The battery is automatically charged when attached to an external power source,
whether the ventilator is operating, in Standby mode, or switched off.
Batteries should undergo a full recharge procedure:
ï‚· Prior to initial use
ï‚· After prolonged storage
ï‚· Every 3 months during normal use
ï‚· If after 8 hours charging, the battery indicator fails to indicate a full charge
Indicator Colors Description Remarks
Green
Battery is fully charged
Gray/Green
Battery partly discharged
Indicator moves
downward in 10%
increments
Gray / Red
Low battery
At 10%, the indicator
changes to red
Gray
Empty battery
Below 7%, the indicator
changes to gray
CAUTION The battery for the iVent
TM
201 is a critical component. It should be
replaced only with batteries supplied by VersaMed. If the user has
been qualified by VersaMed to service the iVent
TM
201, the battery may

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE iVent 201 Series and is the answer not in the manual?

GE iVent 201 Series Specifications

General IconGeneral
BrandGE
ModeliVent 201 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals