SPECIFICATIONS
MONITORED DATA RANGE, RESOLUTION AND
ACCURACY
The following section describes the various specifications and parameters of the
iVent
TM
201.
Parameter Range, resolution, accuracy
Breath type
Range: Type: - Control, assist, patient, sigh
Resolution: N/A
Accuracy: N/A
Respiratory Rate
Range: 0 to 150/min
Resolution: 1/min
Accuracy: 1 for 1 to 12/min
2 for 12 to 150/min
Exhale tidal volume
Range: 0 to 5000 mL
Resolution: 1 mL
Accuracy: 10 mL or 15% (whichever is
greater)
for 0 to 300 mL
10% for 300 to 5000 mL
Exhaled Minute Volume
Range: 0 to 99.9L
Resolution: 0.1 L
Accuracy: 10%
Peak flow
Range: 0 to 140 L/min
Resolution: 1 L/min
Accuracy: 10%
PIP (Peak Inspiratory
Pressure)
Range: 0 to 99 cmH
2
O
Resolution: 1 cmH
2
O
Accuracy:  2 (+ 5 % of reading) cm H2O
Inspiratory time
Range: 0 to 3.0 sec
Resolution: 0.1 sec
Accuracy: 10%
I:E ratio
Range: 1:11 to 3:1
Resolution: 0.1 for 1:1 to 1:5
1 for 1:5 to 1:11 and 3:1 to 1:1
Accuracy: 0.1
Delivered O2%
Range: 21 to 100%