14
Resolution: 1%
Accuracy: 5%
Flow Leak
Range: 0 to 100%
Resolution: 1%
Accuracy: 15%
Mean Airway Pressure
Range: 0 to 99 cmH
2
O
Resolution: 1 cmH
2
O
Accuracy:  2 (+ 5 % of reading) cmH
2
O
Resistance (dynamic)
Range: 0 to 99.9 cmH
2
O/L/s
Resolution: 0.1 cmH
2
O/L/s
Accuracy:  (2 + 20% of actual value)
cmH
2
O/L/s
Compliance (dynamic)
Range: 0 to 99.9 ml/cmH
2
O
Resolution: 0.1 ml/cmH
2
O
Accuracy:  (2 +20% of actual value)
ml/cmH
2
O
RR/Vt-Rapid Shallow
Breathing Index (RSBI)
Range: 0 –200 /min*L
Resolution: 1 /min*L
Accuracy:  (5 + 20% of actual value). /min*L
Static compliance
Range: 0 to 99 ml/cmH
2
O
Resolution: 1 ml/cmH
2
O
Accuracy:  (2 + 20 % of actual value).
ml/cmH2O
Auto PEEP
Range: 0 to 99 cmH
2
O
Resolution: 1 cmH
2
O
Accuracy:  2 (+ 5 % of reading) cmH
2
O
Plateau Pressure
Range: 0 to 99 cmH
2
0
Resolution: 1 cmH
2
0
Accuracy: +2(+5% of reading) cmH
2
0
Time Constant
Range: 0.0 to 9.9 sec
Resolution: 0.1 sec
Accuracy: 20%