36 
 
be substituted and programmed according to the VersaMed Service 
Manual and only with the battery specified in those notes 
 
Full Recharge Procedure 
CAUTION 
Low Battery” or “Empty Battery” alarm appears, the internal batter
 must 
 recharged, as described belo
 
 
Continued usage of the battery after the “Empty Battery” alarm appears may 
disable the battery’s charging capability and/or lead to battery failure. Ventilation 
parameters may not be met in this condition. 
 
NOTE  The full recharge procedure first empties the battery of all its charge.  
Charging it to capacity then allows for proper calibration of the battery 
gauge. 
 
To perform a full recharge procedure: 
1.  Plug the AC power cord into the ventilator.  Verify that the amber Charge LED 
is lit. 
2.  Allow the unit to charge for at least 10 hours. 
 
NOTE  If the Extended battery is in use, charge the unit for at least 24 hours 
 
3.  Switch on the ventilator.  When the Weight Selection window appears, select 
the 70kg patient weight setting. 
4.  Set the pressure alarm to 60 (cmH20). 
5.  Connect the ventilator with a patient circuit to the Rp20 resistor and test 
lung. 
6.  Press Start on the ventilator (See page 59) for directions on starting the 
iVent
TM
201.) 
7.  Adjust the tidal volume (Vt) so that a PIP (peak inspiratory pressure) of 40 is 
attained on each breath – for example, a value of 850. 
8.  Disconnect the power cord.  The AC Disconnect alarm sounds and a pop-up 
window appears.  Press the red Silence button on the keypad to 
remove the pop-up window and silence the alarm. 
9.  Allow the ventilator to run off the battery continuously.  When the Empty 
Battery alarm sounds, clear it by pressing the red Silence button on the 
keypad. 
10. Place the unit in Standby mode (Page 61), then switch it off. Connect the AC 
power cord. 
11. Let the unit charge for at least 10 hours.