GE MEDICAL SYSTEMS
DIRECTION 2300000, REVISION 2 LOGIQ™5 SERVICE MANUAL
8-140 Section 8-8 - PC Block
8-8-2 VIC Assy
Purpose: This is a description on how to remove and replace the VIC assy in the BEP Assy.
8-8-2-1 Tools
Common pillips screwdrivers
8-8-2-2 Needed Manpower
1 persons, 15minutes + travel
8-8-2-3 Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 3.
Left Cover should be removed before proceed. Refer to section 8-5-1 on page 39.
8-8-2-4 Removal Procedure
8-8-2-4-1 Mode setting for VIC assy
VIC assy have two jumpers for power control functionality. The cap position depend on BEP assy.
Before replace the VIC assy, this mode jumper shoub be set up.
Figure 8-160 Mode set for VIC assy