EasyManuals Logo

GE LOGIQ 5 Service Manual

GE LOGIQ 5
452 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #359 background imageLoading...
Page #359 background image
GE MEDICAL SYSTEMS
D
IRECTION 2300000, REVISION 2 LOGIQ™5 SERVICE MANUAL
Chapter 8 Replacement Procedures 8-163
8-9-5 AC Power Box
Purpose: This is a description on how to remove and replace the AC Power Box.
8-9-5-1 Tools
Common pillips screwdrivers
Allen/Unbraco wrench
8-9-5-2 Needed Manpower
1 person, 15 minutes + travel
8-9-5-3 Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 3.
AC Power Assy should be removed before proceed. Refer to AC POWER ASSY DISASSEMBLY
on section 8-9-1 on page 153.
8-9-5-4 Removal Procedure
1.) Unscrew six screws (1-6) and remove the AC Power Box cover. Refer to Figure 8-193.
2.) Remove the AC Power left side cover.
CAUTION
Do not wear the ESD wrist strap when you remove a part of power supply unit. Turn OFF power
and unplug the power cord before removing a part of power supply unit. However be sure to turn
off power and wear the strap before you remove a circuit boards.
Figure 8-193 Unscrew 6 screws to remove AC Power Box
AC Power Box Cover

Table of Contents