GE MEDICAL SYSTEMS
DIRECTION 2300000, REVISION 2 LOGIQ™5 SERVICE MANUAL
8-172 Section 8-10 - Software Loading Procedure
8-10-5-4 Operator Login Password Back-up
The Operator Login Password will be deleted. So, write down the password if necessary.
8-10-5-5 Image Management Guide
NOTE: Before upgrading to R4.x.x software, performe a recommended image backup procedure.
Save As to View on any PC - Use this to save images (Dicom or Raw Dicom) in a computer-friendly
format (.avi or .jpeg) so you can view it on any PC.
Export/Import Data/Images Between Systems - Use this to copy both patient data and images for
specified patient(s) from one system to another.
Save As (Saving Images to CD-ROM to View on Any PC)
To save images to the CD-ROM:
1.) Insert the CD.
2.) Go to Utility --> Connectivity --> Tools(Removable Media instead of Tools in case of BTO5 SW)
Select the Media as CD Rewritable.
3.) Press Format to format the CD-ROM.
4.) Select the image(s) to be saved to CD-ROM, press Menu, and select Save As.
Figure 8-202 Operator Login
Parameters Descriptions
Operator Login Password

Table of Contents