GE MEDICAL SYSTEMS
DIRECTION 2300000, REVISION 2 LOGIQ™5 SERVICE MANUAL
8-196 Section 8-11 - Installing Base Software Load Image
8-11-2-1 Installig Application Software(R4.x.x) (cont’d)
8.) After the installation is completed the system will power off automatically.
NOTE: Do not operate with using mouse or keyboard during installation procedures.
9.) Turn on the system and immediately eject the button on the CD -R drive to eject the CDROM
10.)Verify that the system boots up no error.(It will take a few minutes)