System Setup
3-52 LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Patient Preparation
The ECG traces quality depends very much on the stability and
conductivity of the electrodes during the test, especially during
high stages when the patient movements can cause artifacts.
Here are some basic rules to ensure good electrical contact:
• Shave hair at the electrode contact points.
• Use a special shirt that attached the electrodes and lead
wires to the body.
• Use high quality liquid gel electrodes.
• Make sure that the lead wires do not swing.
Detachable 3 leads each 100cm with clip connectors on the
patient end designed for use with disposable ECG electrodes.
Furthermore, equip 3 repeatable limbs electrode clamps which
can connect with clip connector. Limbs electrodes are white, red
and green (AHA\IEC).
Technical Specification of ECG
Table 3-18: ECG Lead Placement
Lead
Patient Cable Marking
Position on PatientAHA IEC
I RA (White) R (Red) Right Arm
II LA (Black) L (Yellow) Left Arm
III RL (Green) N (Black) Right Leg
Table 3-19: ECG Technical Specifications
Feature Specification
ECG
Patient Cable 3 leads conform to AAMI
Defibrillator Protection: Protected against 360 J discharge
Patient leakage current < 10 uA
Ground Isolation 4 kV rms
Input Impedance >100 Mohm
CMMR > 100 dB
DC max.input +/- 300mv
Frequency range (-3db) 0.05 - 150 Hz

Table of Contents

Other manuals for GE LOGIQ e

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ e and is the answer not in the manual?

GE LOGIQ e Specifications

General IconGeneral
BrandGE
ModelLOGIQ e
CategoryMedical Equipment
LanguageEnglish

Related product manuals