General requirements
LOGIQ e – Basic Service Manual 2-3
5461614-100 English Rev. 6
General requirements
Contents in this Section
• ‘Ultrasound system environmental requirements’ on
page 2-3
• ‘Electrical requirements’ on page 2-6
• ‘Electrical requirements for Docking Cart’ on page 2-8
• ‘EMI limitations’ on page 2-9
• ‘EMI prevention/abatement’ on page 2-10
• ‘Probes environmental requirements’ on page 2-11
Ultrasound system environmental requirements
If the Ultrasound system is very cold or hot
When unpacking the Ultrasound system, allow the temperature
of the Ultrasound system to stabilize before powering up. The
following table describes guidelines for reaching operational
temperatures from storage or transport temperatures.
CAUTION
If the Ultrasound system is very cold or hot, do not turn on its
power until it has had a chance to acclimate to its operating
environment.
Table 2-1: System Acclimation Time Chart
Degree C 50 45 40 35 30 25 20 15 10 5 0 -5
Degree F 122113104958677685950413223
hours 420000000246