Site Preparations
2-4 LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Environmental specifications
The system and Docking Cart should be operated, stored, or
transported within the parameters outlined below. Either its
operational environment must be constantly maintained or the
unit must be turned off.
Table 2-2: System Environmental Requirements
Operational
(with probe)
Storage
LOGIQ e
Transport
LOGIQ e
Temperature 10 - 40°C
50 - 104°F
-5 - 50°C
23 - 122°F
-5 - 50°C
23 - 122°F
Humidity 30 - 75%
non-condensing
10 - 90%
non-condensing
10 - 90%
non-condensing
Pressure 700 - 1060hPa 700 - 1060hPa 700 - 1060hPa
Table 2-3: Environmental Requirements for Docking Cart
Operation Storage Transport
Temperature 10 - 40°C
50 - 104°F
-5 - 50°C
23 - 122°F
-5 - 50°C
23 - 122°F
Humidity 30 - 75%
non-condensing
10 - 90%
non-condensing
10 - 90%
non-condensing
Pressure 700 - 1060hPa 700 - 1060hPa 700 - 1060hPa