Care and Maintenance
10-26 LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Electrical safety tests
Contents in This Section
‘Safety test overview’ on page 10-26
‘Leakage current limits’ on page 10-29
‘Outlet test - wiring arrangement’ on page 10-31
‘Grounding continuity’ on page 10-32
‘Chassis leakage current test’ on page 10-33
‘Data Sheet for enclosure Source Leakage Current’ on
page 10-34
‘Probe leakage current test’ on page 10-35
‘Definition’ on page 10-35
‘Tools’ on page 10-35
‘Generic procedure on probe leakage current’ on
page 10-36
‘Meter Procedure Using Probe Adapter’ on page 10-36
‘No Meter Procedure Using Probe Adapter’ on page 10-36
‘Data Sheet for Transducer Source Leakage Current’ on
page 10-37
Safety test overview
The electrical safety tests in this section are based on and
conform to IEC 60601-1 Medical Equipment Safety Standards.
They are intended for the electrical safety evaluation of
cord-connected, electrically operated, patient care equipment. If
additional information is needed, refer to the IEC 60601-1
documents

Table of Contents