Docking Cart Servicing
12-34 LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Isolated Patient Lead (Sink) Leakage-Isolation Test (continued)
Probe Leakage Current Test
Please refer to ‘Probe leakage current test’ on page 10-35 for
the information.
When There’s Too Much Leakage Current...
Please refer to ‘When there's too much leakage current …’ on
page 10-38 for the information.
Table 12-20:
ECG
Power
Tester
Polarity
Switch
Tester
Ground
Switch
Tester Lead Selector
RL RA LA LL C
ON NORM CLOSED
ON REVERSE CLOSED
ON NORM OPEN
ON REVERSE OPEN
OFF NORM CLOSED
OFF REVERSE CLOSED
OFF NORM OPEN
OFF REVERSE OPEN