Preparing for setup
LOGIQ e – Basic Service Manual 3-11
5461614-100 English Rev. 6
Preparing for setup
Verify customer order
Compare items received by the customer to that which is listed
on the delivery order. Report any items that are missing, back
ordered, or damaged.
Physical inspection
System Voltage Settings
Verify that the system arrived intact (visual inspection).
If the system has been damaged, please refer to ‘Damage in
transportation’ on page i-11 in the beginning of this manual.
• Verify that the scanner is set to the correct voltage. The
Voltage settings for the LOGIQ e Scanner is found on a
label located on the AC adapter.
• 220-240VAC(China); 100-120VAC(USA/Japan);
220-240VAC(Europe, Latin America).
EMI protection
The LOGIQ e has been designed to minimize the effects of
Electro-Magnetic Interference (EMI). Many of the covers,
shields, and screws are provided primarily to protect the system
from image artifacts caused by this interference. For this reason,
it is imperative that all covers and hardware are installed and
secured before the unit is put into operation.
See ‘EMI limitations’ on page 2-9 for more information about
EMI protection.
WARNING
Connecting a LOGIQ e system to the wrong voltage level
will most likely destroy the scanner.

Table of Contents

Other manuals for GE LOGIQ e

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ e and is the answer not in the manual?

GE LOGIQ e Specifications

General IconGeneral
BrandGE
ModelLOGIQ e
CategoryMedical Equipment
LanguageEnglish

Related product manuals