Site Preparations
2-8 LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Electrical requirements for Docking Cart
Table 2-6: Electrical requirements for Docking Cart
PARAMETER AREA LIMITS
Voltage Range 100-240V~ 350VA
Power All applications More than or equal to 750 VA
Line Frequency All applications 50/60Hz (±2Hz)
Power Transients All applications Less than 25% of nominal peak voltage for less
than 1 millisecond for any type of transient,
including line frequency, synchronous,
asynchronous, or aperiodic transients.
Decaying Oscillation All applications Less than 15% of peak voltage for less than 1
millisecond.