General Procedures and Functional Checks
4-12 LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Power shut down (continued)
The SYSTEM - EXIT menu, used when switching off the unit,
gives you these choices:
• Logoff
Use this button to log off the current user.
The system remains ON and ready for a new user to log on.
If the Logoff button is dimmed, it indicates that no user is
logged on to the unit at the moment.
• Shutdown
Use this button to shut down the system. The entire system
will shut down. It is recommended to perform a full shutdown
at least once a week.
• Sleep Mode
Use this button when you do a portable exam in order to
reduce the time to start up the system.
• Cancel
Use this button to exit from the System-Exit menu and
return to the previous operation.
• Exit
(Only available when logged in as GE Service with Service
Dongle)
Select this button when you want to exit to the Windows
Desktop.
NOTE: If you need to restart LOGIQ e when logged on to the
Windows Desktop, ensure that you do a complete power
down (Shut Down). This is required to power up the Front
End Processor.
System shutdown
Disconnect the Mains Power Cable is necessary. For example:
Relocating the scanner.
CAUTION
DO NOT unplug and/or transport the unit until after the power
off sequence has been completed. Failure to do so may result
in corrupted patient files.