Care and Maintenance
10-4 LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Protecting Health Information
Prevent from copying data including Patient Information to
external storage device.
Select Utility -> Admin -> System Admin and check the box in
the portion of Protecting Health Information (PHI). Select
Save and reboot the system.
Figure 10-2. Protecting Health Information
Prevent writing patient information to external storage:
Media: USB Memory / SD Card / USB HDD / CD-R / DVD-R
Function: Export / MPEGVue / EZBackup / EZMove / Save As /
Save As Images / Report Save As / Backup Patient Archive and
Report Archive from Utility.