Docking Cart Setup
11-6 LOGIQ e – Basic Service Manual
5461614-100 English Rev. 6
Setup Warnings (continued)
• After being transported, the unit may be very cold or hot. If
this is the case, allow the unit to acclimate before you turn it
on. It requires one hour for each 2.5°C increment it's
temperature is below 10°C or above 30°C.
CAUTION
Equipment damage possibility. Turning the system on without
acclimation after arriving at site may cause the system to be
damaged.
Table 11-2: Acclimation Time
°C6055504540353025201510
°F 140 131 122 113 104 95 86 77 68 59 50
hrs86420000000
°C 5 0 -5 -10 -15 -20 -25 -30 -35 -40
°F413223145 -4 -13-22-31-40
hrs2468101214161820