om 5184516-100 Rev. 5 11-4
11-4 Generator Ouput Power:
TABLE 11-5
JEDI GENERATOR POWER
The JEDI can deliver the power specified in this output for at least 100ms,
depending on the X–ray tube used. Refer to the tube rating charts supplied with
the x–ray tube.
In accordance with IEC requirements, TABLE 11-7 defines the reference Current
Time Product.
TABLE 11-6
NOMINAL ELECTRIC POWER (1000KV–0.1S):
TABLE 11-7
PARAMETER SELECTION RANGES OF CONSOLE READOUTS
kV 65kW 80kW
40–80 800 mA 1000 mA
100 630 mA 800 mA
125 500 mA 630 mA
150 400 mA 500 mA
kW
k
65 80
I (max)
mA
630 800
All Modes
Selectable kV 40–150
Decimal scale steps of 1kV Jump of 10kV
Renard scale 40, 41, 42, 44, 46, 48, 50,
52, 55, 57, 60, 63, 66, 70,
73, 77, 81, 85, 90, 96, 102,
109, 117, 125, 133, 141,
150
The next kV displayed is
the third Renard step value
following the initial display
value of kV.