om 5184516-100 Rev. 5 11-5
TABLE 11-8
ACCURACY READINGS ON CONSOLE READOUTS
Constancy of AIR KERMA
Nominal Shortest Irradiation Time:
80kV, 10ms – without correction
80kV, 7ms – with correction by HV Generator.
Mode : 3 point
Selectable Resultant
mA 26% ms 26A% mAs 26% increments
Non–AEC
AEC
0.5 thru 630 mA
32 steps
0.5 thru 630 mA
32 steps
JEDI 65 kW: 10 thru
800 mA/20 steps
1 thru 630ms 39 steps
JEDI 80 kW: 10 thru
1000 mA/21 steps
1 thru 630ms 39 steps
Auto Exposure Control: AEC
AEC Technique Sensing Device Sensing area Manual density
correction
Post–display: Auto-
matic for 5 sec.
AEC Cassette Ionization
Chamber: up to 2
3 rectangular
cells
5 steps, –2thru
+2, by 26%
increment
mAs, ms and mA
Mode: 2 point
User Selectable
MAs 26% increments
Non–AEC 0.5 thru 630mAs 32 steps
AEC 0.5 thru 500mAs 31 steps
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).