om 5184516-100 Rev. 5 11-6
TABLE 11-9
ACCURACY OF CONSOLE READOUTS
NOTE: In RAD mode, kVp accuracy is ensured after 10ms.
Display before exposure
Conditions parameter mode Basic Accuracy
Radiography
Exposure
Time
<= 40
kVp all +/– (3%, +2 kVp)
mA all +/– (10%, +1 mA)
ms 3 point timer +/–(5%, +0.5ms)
ms AEC (2/3 point) +/– 5%
mAs AEC (2/3 point) +/– 10%
mA AEC (3 point) +/– 15%
Radiography
Exposure
Time
> 40 ms
kVp all +/– (3%, +2 kVp)
mA all +/– (5%, +1 mA)
ms 3 point timer +/–(5%, +0.5ms)
ms AEC (2/3 point) +/– 5%
mAs AEC (2/3 point) +/–(5%, +0.1 mAs)
mA AEC (3 point) +/– 10%
FOR TRAINING PURPOSES ONLY!