om 5184516-100 Rev. 5 B-7
7. When the “CONTINUE” button is pressed, the screen will transition to the
Step 2 Instruction Screen with the following instructions:
Step 2 of 3
• Adjust the tower so the the SID is equal to the value displayed in the
screen instructions, +/-5 mm. Lock the tower at that height.
• Place a 12x12x1/32 inch copper sheet on the top of the table in the center
of the field of view. Ensure that the copper sheet attenuates the entire X-
ray field.
• Depress the hand switch and release after at least 2 images have been
acquired. (The number of images acquired is displayed on the image
monitor.)
• Select the "CONTINUE" button when finished.