DEFAULT SETTINGS
When the Restore Defaults option is selected, the settings will be changed to the factory defaults as specified in the table
below.
NOTE:
Change power-up default in Menu, do not change on display.
Item Factory Default
Display: Channel 1 Source ECG II
Display: Channel 1 Gain 1x
Display: Channel 2 Source SpO
2
Display: Channel 2 Gain 1x
ECG Info Box: Low Alarm 40
ECG Info Box: High Alarm 180
SpO2 Info Box: Low Alarm Off
SpO2 Info Box: High Alarm Off
Pacing Info Box: On/Off/Pause Off
Alarm Button: Off
Menu->Settings->Defibrillation->Manual Mode 150J
Menu->Settings->Defibrillation->Use Auto Sequence Yes
Menu->Settings->Defibrillation->Auto Sequence Energy #1 200
Menu->Settings->Defibrillation->Auto Sequence Energy #2 200
Menu->Settings->Defibrillation->Auto Sequence Energy #3 270
Menu->Settings->Defibrillation->Detection 160
Menu->Settings->Defibrillation->SVT 240
Menu->Settings->Defibrillation->Use SVT For Analysis Yes
Menu->Settings->Pacing->Mode Demand
Menu->Settings->Pacing->Rate 60
Menu->Settings->Channel Display->Channel 1 Source ECG II
Menu->Settings->Channel Display->Channel 1 Gain 1x
Menu->Settings->Channel Display->Channel 2 Source SpO2
Menu->Settings->Channel Display->Channel 2 Gain 1x
Menu->Settings->Channel Display->Line Filter Off
Menu->Settings->Channel Display->Muscle Filter On
Menu->Settings->Alarms and Sound->Audible Alarm Off
Menu->Settings->Alarms and Sound->Volume Loud
Menu->Settings->Alarms and Sound->Pulse Beep Soft
Menu->Settings->Alarms and Sound->Low Heart Rate 40
Menu->Settings->Alarms and Sound->High Heart Rate 180
Menu->Settings->Alarms and Sound->Low SpO2 Sat Off
Menu->Settings->Alarms and Sound->High SpO2 Sat Off
Menu->Settings->System Settings->Display Settings->Level Bright
Menu->Settings->System Settings->Printer Settings->Duration 20 Sec
Menu->Settings->System Settings->Printer Settings->Events Shock 20 Seconds
Menu->Settings->System Settings->Power-Up Settings->Mode Manual Mode
Menu->Settings->Date and Time->Hour N/A
Menu->Settings->Date and Time->Minute N/A
Menu->Settings->Date and Time->Format 24 Hour
Menu->Settings->Date and Time->Day N/A
Menu->Settings->Date and Time->Month N/A
Menu->Settings->Date and Time->Year N/A
Menu->Settings->Date and Time->Format Day.Month.Year
Menu->Settings->User Settings->Facility Name <blank>
Menu->Settings->User Settings->Set Password <blank>

Table of Contents

Other manuals for GE Responder 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Responder 2000 and is the answer not in the manual?

GE Responder 2000 Specifications

General IconGeneral
BrandGE
ModelResponder 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals