2026116-001 Revision B Responder™ 2000 Page 120
SPECIFICATIONS
DISPLAY
Size: 115.2mm X 86.4mm
Type: Transmissive Color TFT LCD
Resolution: 320 X 240 pixels
Number of waveform channels: 2
DEFIBRILLATOR
Waveform: Biphasic truncated exponential
Charge Time: 7 seconds nominal
Delivery Method: Via pads, paddles or spoons
Paddles and Pads Energy Selections (50
load):
2, 3, 5, 7, 10, 15, 20, 30, 50, 70, 100, 150, 200, 270
Internal Paddle Energy Selections (50 load): 2, 3, 5, 7, 10, 15, 20, 30, 50
Energy Accuracy: ± 15% onto a 50 load or ± 3 Joules, whichever is greater
NON-INVASIVE PACING
Output Waveform: Rectilinear, constant current
Delivery Method Via Pads
Pulse Width: 40 milliseconds ±4mS
Pacing Modes: Demand or Asynchronous (Fixed Rate)
Pacing Rate: Operator adjustable: 30 to 180 ppm, ± 5%
Pacing Current: 0mA to 140mA
SPO
2
Display
Plethysmogram, digital value of percent saturation, and upper and lower
alarm limits.
Display Update Period
8 seconds or less
Saturation Range
0 to 100%, in 1% increments
Low Saturation Alarm Off, 81-98% in 1% increments
High Saturation Alarm
Off, 95-100% in 1% increments
Pulse Rate Range
20 to 255 BPM, in 1 BPM increments
Audible Alarm Delay 10 seconds or less
Visual Alarm Delay 2 seconds or less
Pulse Rate Accuracy
30 to 250 bpm: ± 2 digits or ± 2%, whichever is greater (without motion)
30 to 250 bpm: ± 3 digits (during low perfusion)
Saturation Accuracy +/- 2 digits from 70% SpO2 to 100% SpO2 with D-O probes, except +/- 3
digits for D-O Ear Probe. OEM board accuracy +/-3 digits from 70%