2026116-001 Revision B Responder™ 2000 Page 88
TO CHANGE SETTINGS FROM STARTUP MENU
1. From the startup menu, use Rotary Selector Knob to move focus to Pacing mode box in the bottom right hand side of
the menu.
2. Press the Rotary Selector knob to change the Pacing Settings.
3. After selecting the desired pacing mode, current, and rate settings, turn the selector knob to highlight the Pacing
mode box and press the rotary knob to confirm the settings.
4. Use Rotary Selector Knob to move focus to Pacing Mode Edit box and press Rotary Selector knob to select it.
PACING DEFAULT PARAMETERS
Mode Current Rate (PPM)
Factory Default: Demand
Available Settings: Demand, Fixed
Factory Default: 0 mA
Available Settings: 0 mA – 140mA
(increments of 5)
Factory Default: 60
Available Settings: 30 - 180
(increments of 5)