2026116-001 Revision B Responder™ 2000 Page 86
DEFAULT SHOCK ENERGY SETTINGS
The Default Shock Energy Settings menu allows default settings to be selected for Manual Mode, Auto Sequence
Mode and Use Auto Sequence.
Manual Mode Auto Sequence Mode Use Auto Sequence Mode
Factory Default: 150 Joules
Available Settings: 2, 3, 5, 7,10,15,
20, 30, 50, 70, 100, 150, 200, 270
Joules
Factory Default: 200, 200, 270 Joules
Available Settings for first, second and
third shock: 2, 3, 5, 7, 10, 15, 20, 30,
50, 70, 100, 150, 200, 270 Joules
Factory Default: Yes
Available Settings: Yes or No
NOTE: The maximum energy for spoons is 50 joules.
FIBRILLATION ANALYSIS THRESHOLDS
The Fibrillation Analysis Thresholds menu allows default settings to be selected for Detection (BPM), SVT (BPM), and
Use SVT for analysis.
Detection (BPM) SVT (BPM) Use SVT for analysis
Factory Default: 160
Available Settings: 120 through 240 in
(increments of 5)
Factory Default: 240
Available Settings: 120 through 240
(increments of 5)
Factory Default: Yes
Available Settings: Yes or No