2026116-001 Revision B Responder™ 2000 Page 109
VISUAL INSPECTION
The Responder 2000 and its accessories should be carefully inspected prior to installation, daily, and each time the
equipment is serviced.
• Carefully inspect the equipment for physical damage
• Inspect all external connections for loose connectors or frayed cables.
• Inspect the graphics display for marks, scratches, or other damage.
• Verify that the Safety label on back of the Responder 2000 is clearly legible
INSTRUCTION INSPECT FOR RECOMMENDED REMEDY
Examine the case connectors and
accessories
Foreign substances
Clean the Responder 2000 and its
accessories as described.
Damage or cracks Replace if damaged or cracked
Cables or Connector Replace if damaged or cracked
Pins bent or discolored
With unit plugged into an active
AC outlet, turn power on
Battery Indicator is showing battery
low
Plug Responder 2000 in using power
cord to charge battery.
Examine accessory cables Foreign substances Clean the cables as described in
the Section 5
Broken parts, cracks, damage, Replace cable if any abnormalities
or extreme wear, broken or bent are found.
connectors and pins, after bending
and flexing the cable
Examine disposable accessories Expired ECG or Responder 2000 Replace any products approaching
pads or past their expiration dates.
WARNING: After the visual inspection, if the Responder 2000 and/or its accessories are damaged please contact
Customer Service. The Responder 2000 will need to be repaired. The accessories should be disposed of appropriately
and replacement parts shall be ordered.