2026116-001 Revision B Responder™ 2000 Page 35
Z-BAR™ INDICATOR
The Z-Bar provides a relative visual graphical indicator of the total transthoracic impedance between the two defibrillation
pads or paddles. The Z-Bar is used in the assessment of:
Adequate Pad, Paddle , or Spoon placement
Pad or Paddle quality and integrity
Pad or Paddle adhesion to the patient’s skin
Pad or Paddle connection to the Responder 2000
Provides for quick assessment between OFF and SHORTED
Z-BAR FOR PADS AND PADDLES
SECTION MEASURED IMPEDANCE
RANGE (OHMS)
DESCRIPTION COLOR FILL
1 0-24
Lower Limit – Non-operational range Red
2 25-35 Lower marginal operating range. Indicates
potential degradation in quality or position
Yellow
3 36-135 Normal operating range Green
4 136-200 (for pads)
>136 (for paddles)
Upper marginal operating range. Indicates
potential degradation in quality or position
Yellow
5 >201 (for pads) Upper Limit– Non-operational range Red
Z-BAR FOR SPOONS
SECTION MEASURED IMPEDANCE
RANGE (OHMS)
DESCRIPTION COLOR FILL
1 0-9 Lower Limit – Non-operational range Red
2 10-15 Lower marginal operating range. Indicates
potential degradation in quality or position
Yellow
3 16-75 Normal operating range Green
4 76-200 Upper marginal operating range. Indicates
potential degradation in quality or position
Yellow
5 >201 Upper Limit– Non-operational range Red