2026116-001 Revision B Responder™ 2000 Page 37
CHARGE BUTTON
The charge button is used to manually charge the Responder 2000 to the selected energy level. This button is only used
in Manual Mode. This button is disabled when paddles are connected to the Responder 2000. In this case, the
Responder 2000 will be charged only from the paddle charge button. The button will also be disabled when the Z-Bar is
in the red range for pads or spoons.
SHOCK BUTTON
The shock button activates (flashes red) only when the system is charged and ready to deliver therapy to the patient.
Press and hold the shock button until therapy is delivered. This button is disabled when the paddles are connected to the
Responder 2000. If pads or spoons are used, the shock button will only be activated with good impedance.