EN 60601-1-2 COMPLIANCE
WARNING: RF Interference
Known RF sources, such as cell phones, radio or TV stations, and two-way radios, may cause unexpected
or adverse operation of this device. Consult qualified personnel regarding system configuration.
WARNING: Equipment Configuration
The equipment or system should not be used adjacent to, or stacked with other equipment. If adjacent or stacked use is
necessary, test the equipment or system to verify normal operation. Refer to the Electromagnetic Immunity information in
the appendix.
COMPLIANT CABLES AND ACCESSORIES
WARNING:
The use of accessories, transducers and cables other than those specified in Section 7 of this manual may increase
emissions or decrease immunity performance of the device/system.
The table below lists cables, transducers, and other applicable accessories with which GE Medical Systems claims EMC
compliance.
NOTE: Any supplied accessories that do not affect EMC compliance are not included.
Part Number Description Maximum Lengths
ECG Cables and Lead wire set
2017003-001 IEC 5 lead ECG cable 3.6 m / 12 ft
2017003-003 AHA 5 lead ECG cable 3.6 m / 12 ft
414556-001 5 lead wire set AHA 0.74m / 29 inch
414556-003 5 lead wire set IEC 0.74m / 29 inch
2021141-001 3 lead wire set IEC , combined cable and lead wire 4.8 m / 16 ft
2021141-002 3 lead wire set AHA , combined cable and lead wire 4.8 m / 16 ft
Defibrillator Cables
2030247-001 Cable for defibrillator electrode pads 4 m / 13 ft
2030249-001 Cable for defibrillator contact paddle, with handle 4 m / 13 ft
SPO
2
Cables and Sensors
OXY-ES3 OxyTip+® interconnect cable 3 m / 10 ft
OXY-F-UN Finger Sensor 1 m / 3.3 ft
OXY-E-UN Ear Sensor 1 m / 3.3 ft
OXY-W-UN Wrap Sensor 1 m / 3.3 ft
Power cords
2019204-007 Power cord, North America 2.5 m / 8 ft
2020387-002 Power cord, Europe 2.5 m / 8 ft