2026116-001 Revision B Responder™ 2000 Page 112
EXTERNAL COUNTER ELECTRODE FOR INTERNAL DEFIBRILLATION
Disconnect the electrode from its lead before cleaning or sterilizing it.
Clean the electrode by rubbing it down with a cloth moistened with soap water. Use a disinfectant for
disinfection. Do not immerse the electrode in the liquid.
Low-temperature plasma sterilization is the recommended sterilization method. Alternative methods are ETO
sterilization or ionizing radiation. (Please note: Frequent ETO sterilization reduces the life expectancy of the
plastic material!) Do not autoclave the electrodes.
CARING FOR RECHARGEABLE BATTERIES
CALIBRATING THE BATTERY FUEL GAUGE
Time, repeated partial charges and discharges, and battery aging will lead to inaccuracy of the battery fuel gauge. This is
corrected by performing a battery calibration cycle. To maintain useful battery life, it is recommended to calibrate the
rechargeable battery:
Once every two months
When the run time of the battery becomes noticeably shorter
When the predicted run times become noticeably inaccurate
To calibrate a Responder 2000 battery, see Section 2.
RECYCLING THE BATTERIES
When the battery no longer holds a charge, it should be replaced. This condition may occur after 2.5 years or 300
charge/discharge cycles. The batteries are recyclable. Remove the old battery from the Responder 2000 and follow your
local recycling guidelines.

Table of Contents

Other manuals for GE Responder 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Responder 2000 and is the answer not in the manual?

GE Responder 2000 Specifications

General IconGeneral
BrandGE
ModelResponder 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals