2026116-001 Revision B Responder™ 2000 Page 69
TO USE PACING MODE
To Enable Pacing:
1. Apply ECG electrodes to patient.
2. Connect ECG electrodes to the device.
3. Turn on the Responder 2000.
4. Check ECG and vital signs
5. Attach pace pads to patient.
6. Connect pace pads to the Responder 2000.
7. Use Rotary Selector Knob to move focus Pacing Info box, at the display’s lower right.
8. Select the Pacing Info box. If pacing is not enable, the Pacing Info box is not selectable.
9. Use Rotary Selector Knob to traverse though selectable items.
10. Turning clockwise, the following items should be selectable, in order, wrapping around:
a) Pacing Type (Fixed, Demand), Default is Demand.
b) PPM ( 30-180 PPM, increments of 5), Default is 60 PPM
c) Exit Pacing Info Box (“x” in corner)
d) Pacing On/Off/Pause (Pause is only available when Pacing is turned On)
e) Pacing Current (0mA-140mA, increments of 5). This option is only selectable when pacing is turned on.
11. Turn the Rotary Selector Knob to turn on the pacing.
12. When Pacing is turned on, focus automatically goes to the Pacing Current selection. Set desired pacing current.
13. Turn the rotary knob to select pacing rate and set desired options for pacing rate.
To Stop Pacing:
1. Turn the pacemaker OFF
2. Disconnect pace pads from the Responder 2000.
3. Remove pace pads from patient.
4. If monitoring of the patient is no longer required, turn off the Responder 2000.
5. Remove ECG electrodes from patient.