2026116-001 Revision B Responder™ 2000 Page 122
PHYSICAL DIMENSIONS
Height: 10.8 inches/27.4 cm
Width: 11.7 inches/29.7 cm
Depth: 7.4 inches/18.8 cm
Weight: Less than 10lbs/4.5kg, excluding battery, paddles, and a full roll of paper.
ENVIRONMENTAL REQUIREMENTS
OPERATING CONDITIONS
Temperature: 0°C to 50°C (32 °F to 122°F)
Humidity: 10% to 95% RH, non-condensing
Air Pressure altitude: -500 ft (103kPa) to 15000 ft (57kPa)
STORAGE AND SHIPPING CONDITIONS
Temperature (Responder 2000): -20°C to 60°C (-4°F to 140°F)
Temperature (Pads): -12°C to 43°C (10°F to 110°F) for 2 years.
NOTE: Do not exceed 38°C (100°F) for periods greater than 6 months in duration. Always store the pads in
their pouch to maintain freshness. Do not use if gel has dried out.
Humidity: 10% - 95% RH, non-condensing
Air Pressure altitude: -500 ft (103kPa) to 15000 ft (57kPa)
WARNING:
Electrode performance may be adversely affected by pre-attaching and storing with defibrillator cable or
exposure to air for long periods of time. These electrodes are not recommended for electrosurgery.
PRECAUTION: Environment of use
Responder 2000 is designed for indoor use. Operator must confirm that the environment of use meets the required
operating environmental specifications before using.
PRECAUTION: Cold Environments
If the Responder 2000 is stored in an environment with a temperature below the operating temperature, the unit should be
allowed to warm up to the needed operating temperature before using.