2026116-001 Revision B Responder™ 2000 Page 80
ADJUSTING HEART RATE ALARM LIMITS
1. Rotate Rotary Selector Knob and select Heart Rate Alarm
2. Rotate Rotary Selector Knob and Select Low Alarm Limit Range
Available Low Alarm Settings: Off, 25-120 (in increments of 5)
Default: 40
3. Press Rotary Selector Knob to confirm setting
4. Rotate Rotary Selector Knob and Select High Alarm Limit Range
Available Settings: 40-300 (increments of 5), Off
Default: 180
SETTING THE ECG SOURCE AND GAIN
1. To set the source for the ECG waveforms, rotate the Rotary Selector Knob to highlight the current ECG selection.
2. Press Rotary Selector knob to select setting.
3. Turn the Rotary Selector Knob again to increment the source selection. The options are:
• ECG I, ECG II or ECG III
• aVR
• aVF
• aVL
• V
• Paddles
4. Press the Rotary Selector Knob to confirm your selection.
5. To set the Gain for the displayed waveform, rotate the Rotary Selector knob to highlight the gain value.
NOTE: The gain in display does not affect the gain on the printout. The printout is always at 1x gain.
6. Rotate the Rotary Selector Knob to increment the Gain value
Available options: 0.25x, 0.5x, 1x, 2x, and 4x.
7. Press the Rotary Selector Knob to confirm your choice. The waveform size will adjust accordingly.