2026116-001 Revision B Responder™ 2000 Page 116
8. Monitor Testing
The performance of the monitor can be tested with the 2025269-003 Simulator.
a) Make sure the battery is at least partially charged.
Power the Responder 2000 from the battery only, i.e.
disconnect AC power input to the device.
If the battery is not partially charged,
wait and partially charge the battery.
b) Connect the ECG leads of the Responder 2000 to
the simulator. Turn on the simulator. Select a
normal sinus rhythm on the simulator. Make sure
you are able to select all leads, and that the ECG
signal is present, and is not distorted or noisy. Make
sure there are no false pace pulse detections.
If the Responder 2000 is unable to
turn on, or unable to select all leads,
or the ECG signal is not present, or
is distorted or noisy, contact
Customer Service.
c) Turn the power to the Responder 2000 off. Remove
the battery. Power the Responder 2000 from AC
input only.
No Action
d) Connect the pads cable from the Responder 2000 to
the simulator. Turn on the simulator. Select a
normal sinus rhythm on the simulator. Make sure the
ECG signal is present, and is not distorted or noisy.
Make sure there are no false pace pulse detections.
If the Responder 2000 is unable to
turn on, or if the ECG signal is not
present, or is distorted or noisy,
contact Customer Service.
9. SpO
2
Testing (if Responder 2000 has this option)
Connect desired SpO
2
cable and connector combination
to Responder 2000 and apply sensor to a test person.
No Action
Verify SpO
2
is plausible
If SpO
2
is displayed not plausible,
please contact Customer Service.
10. Pacing
The performance of the pacemaker can be tested with the 2025269-003 Simulator.
Connect the pads cable from the Responder 2000 to the
simulator. Turn on the simulator and the Responder 2000.
Turn on the pacer and select pacer current > 60 mA.
Verify pacer pulse LED at the simulator is flashing in the
rhythm of the pace pulses.
If this is not the case, please contact
Customer Service.