2026116-001 Revision B Responder™ 2000 Page 56
7. Now trigger the shock within 30 seconds. To do so, simultaneously press the buttons (A) and (B) on the paddles.
NOTE: When the patient is a child under 8 years of age or weighs less than 55 lbs (25 kg), the Responder 2000
should be used with Pediatric Defibrillation paddles and pediatric energy protocols. These paddles are
integrated into the standard paddle set and are available by removing the adult contact plates off the paddles.
The paddles have two different contact surfaces; a large one (can be removed) for the defibrillation of adults and a smaller
one for the defibrillation of children.
Remove the large contact surface for pediatric use:
Press on the lock button 1
Slide the contact surface 2 towards the front and take it off the paddle.
When re-installing it, the large contact surface must audibly click into place.