2026116-001 Revision B Responder™ 2000 Page 62
TO USE MANUAL MODE (CONTINUED)
2. After selecting desired energy, press the Rotary Selector Knob.
3. Push the Charge button.
4. Warn bystanders to stand clear of the patient. Ensure that you do not touch the patient.
5. After the Responder 2000 is charged, push and hold the Shock button(s) until therapy is delivered.
6. The message Shock Delivered displays for eight (8) seconds once the shock has been delivered.
7. At any time during the manual defibrillation sequence, the operator can change the energy setting or exit the mode
completely by turning the Rotary Selector Knob and pressing Exit OR by pushing the Manual button.
NOTE: The operator has thirty (30) seconds to deliver therapy before the Responder 2000 disarms and
aborts therapy.
NOTE: Alarm Silence symbol is displayed, indicating that no audible tone warnings will be heard, only
written warning message will be displayed on the graphics display.