2026116-001 Revision B Responder™ 2000 Page 65
TO USE SEMI-AUTO SHOCK MODE (CONTINUED)
1. Press the Rotary Selector Knob to Analyze.
2. If a shockable rhythm is detected, the Responder 2000 will automatically charge to the appropriate energy level. The
sequence starts with the first displayed value then increments through the three increasing values.
3. When the charge is complete, the operator is prompted to push Shock button(s).
NOTE: The operator has thirty (30) seconds to deliver therapy before the Responder 2000 disarms and aborts
therapy.
4. After three (3) shocks in Semi-auto Shock mode, the Responder 2000 continues to use the last delivered energy.