EasyManua.ls Logo

GE Solar 8000M/i - Page 78

GE Solar 8000M/i
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-26 Solar 8000M/i patient monitor 2026265-075C
Maintenance and checkout: Checkout procedure
4. Verify the SRC-MAX default indicators are as follows:
Heart rate = 60 bpm
Light = LOW
MOD = LOW
5. Verify the following SpO
2
readings for saturation and pulse rate:
Saturation (%): 75 ± 2
Rate (bpm): 60 ± 2
6. Press and release the heart rate button on the SRC-MAX. Verify
default indicators are lit as follows:
Heart rate = 200 bpm
Light = LOW
%SpO
2
= 75
MOD = LOW
7. Verify the following SpO
2
readings for saturation and pulse rate:
Saturation (%): 75 ± 2
Rate (bpm): 200 ± 3 (194 to 206)
8. Press and release the light button on the SRC-MAX. Verify default
indicators are lit as follows:
Heart rate = 200 bpm
Light = HIGH
%SpO
2
= 75
MOD = LOW
9. Verify the following SpO
2
readings for saturation and pulse rate:
-Saturation (%): 75 ± 2
-Rate (bpm): 200 ± 3 (194 to 206)
10. Press and release the %SpO
2
button on the SRC-MAX. Verify default
indicators are lit as follows:
Heart rate = 200 bpm
Light = HIGH
%SpO
2
= 90
MOD = LOW
11. Verify the following SpO
2
readings for saturation and pulse rate:
Saturation (%):90 ± 2
Rate (bpm): 200 ± 3 (194 to 206)

Table of Contents

Related product manuals