4-26 Solar 8000M/i patient monitor 2026265-075C
Maintenance and checkout: Checkout procedure
4. Verify the SRC-MAX default indicators are as follows:
 Heart rate = 60 bpm
 Light = LOW
 MOD = LOW
5. Verify the following SpO
2
 readings for saturation and pulse rate:
 Saturation (%): 75 ± 2
 Rate (bpm): 60 ± 2
6. Press and release the heart rate button on the SRC-MAX. Verify 
default indicators are lit as follows:
 Heart rate = 200 bpm
 Light = LOW
 %SpO
2
 = 75
 MOD = LOW
7. Verify the following SpO
2
 readings for saturation and pulse rate:
 Saturation (%): 75 ± 2 
 Rate (bpm): 200 ± 3 (194 to 206)
8. Press and release the light button on the SRC-MAX. Verify default 
indicators are lit as follows:
 Heart rate = 200 bpm
 Light = HIGH
 %SpO
2
 = 75
 MOD = LOW
9. Verify the following SpO
2
 readings for saturation and pulse rate:
 -Saturation (%): 75 ± 2 
 -Rate (bpm): 200 ± 3 (194 to 206)
10. Press and release the %SpO
2
 button on the SRC-MAX. Verify default 
indicators are lit as follows:
 Heart rate = 200 bpm
 Light = HIGH
 %SpO
2
 = 90
 MOD = LOW
11. Verify the following SpO
2
 readings for saturation and pulse rate:
 Saturation (%):90 ± 2 
 Rate (bpm): 200 ± 3 (194 to 206)