2026265-075C Solar 8000M/i patient monitor 4-27
Maintenance and checkout: Checkout procedure
12. Press and release the MOD button on the SRC-MAX. Verify default
indicators are lit as follows:
Heart rate = 200 bpm
Light = HIGH
%SpO
2
= 90
MOD = HIGH
13. Verify the following SpO
2
readings for saturation and pulse rate:
Saturation (%):90 ± 2
Rate (bpm): 200 ± 3 (194 to 206)
SpO2 functions (Nellcor OxiSmart, Tram only)
1. Connect the Nellcor pulse oximeter SRC-2 to the sensor connector.
2. Make sure that the SRC-2 IR and RED LED drive indictors are both
lit.
3. Set the RCAL/MODE switch to RCAL 63/LOCAL.
4. Set the LIGHT and MODULATION switches as follows:
Set the LIGHT switch to HIGH 1.
Set the MODULATION switch to HIGH.
Set the RATE switch to 112.
Allow the Tram a few seconds to obtain a steady reading.
5. Verify the following SpO2 readings for saturation and pulse rate:
Saturation (%): 81 ± 2
Rate (bpm): 112 ± 2% (110 to 114)
6. Set the LIGHT and MODULATION switches as follows:
Set the LIGHT switch to LOW.
Set the MODULATION switch to LOW.
Set the RATE switch to 201.
7. Verify the following SpO2 readings for saturation and pulse rate:
Saturation (%): 81 ± 2
Rate (bpm): 201 ± 3% (195 to 207)
8. Set the LIGHT and MODULATION switches as follows:
Set the LIGHT switch to HIGH 2.
Set the MODULATION switch to LOW.
Set the RATE switch to 38.
9. Verify the following SpO2 readings for saturation and pulse rate:
Saturation (%): 81 ± 2
Rate (bpm): 38 ± 2% (37 to 39)